Field gauge calibration
Permanent outdoor installation
Burst disc testing
Pressure safety valve (PSV) testing
Hydraulic torque wrench indication
Torque datalogging
Pressure regulator testing
|
Pipeline static pressure testing
Chart recorder replacement
Pressure regulator adjustment
Pressure switch testing
Wellhead monitoring
Filter performance testing
Hydrostatic leak testing
|
Included with every new XP2i Digital Pressure Gauge is an A2LA accredited calibration certificate with test data recorded at 5 different temperatures - proving that it meets published specifications over the entire operating range. You won't get this kind of detailed report anywhere else!
Specifications
Specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.
|
Accuracy
20% to 100% of Full Scale: ±(0.1% of Reading).
0 to 20% of Full Scale: ±(0.02% of Full Scale).
Vacuum, for 2MPa (300 PSI, 20 bar, 20kg/cm2) and lower pressure gauges:
0 to -99.9 kPa (-14.5 PSIG): ±(0.25% of Full Scale), where F.S.: -99.9 kPa (-14.5 PSIG).
(Not recommended for continuous use at high vacuum.)
Pressure Ranges
(see table below for details)
Note:
Density of water conversion can be selected via the keypad: 4°C, 60°F or 20°C/68°F.
Temperature
Operating & Compensated: -10°C to 50°C (14°F to 122°F).
Storage: -40°C to 75°C (-40°F to 167°F).
Backlit Display
Description: 5 Full (seven segment) digits.
Display rate: 3 readings per second.
Numeral display height: 16.5mm (0.65") single line display or 14 mm (0.55") for main numerals of dual line display.
Sensor
Description: All welded stainless steel, with a permanent fill diaphragm seal (filled with Dow Corning 200).
Power
Battery: Three size AA (LR6) batteries.
Battery life: 1500 hours typical (alkaline battery).
Enclosure
Description: Aluminum alloy, sealed to IP67 ratings. (Excludes the WT Series)
Weight: 555g (19.6 oz.), including batteries.
|
Intrinsic Safety
Exia Intrinsically Safe, Class I, Division 1, Groups A, B, C & D, Temperature code T4

Other Certifications
DNV classification certifies all materials, components and systems relevant to the safe operation and quality of ships.

Connections

Pressure connection: 1/4 male NPT or G 1/4 B BSP.
Electrical connection: DB9, RS-232 (environmentally sealed).
Media Compatibility
Liquids and gases compatible with 316 stainless steel.
|
Pressure Ranges
P/N
Prefix |
P/N
Prefix |
P/N
Prefix |
Over
pressure |
PSI |
kg/cm2 |
inch Hg |
inch H2O |
mm Hg |
mm H2O |
kPa |
bar |
mbar |
| 15PSI |
1BAR |
100KPA |
6.5x |
0.001 |
0.0001 |
0.001 |
0.01 |
0.01 |
1 |
0.01 |
0.0001 |
0.1 |
| 30PSI |
2BAR |
200KPA |
3.0x |
0.001 |
0.0001 |
0.001 |
0.01 |
0.1 |
1 |
0.01 |
0.0001 |
0.1 |
| 100PSI |
7BAR |
700KPA |
2.0x |
0.01 |
0.0001 |
0.01 |
0.1 |
0.1 |
1 |
0.01 |
0.0001 |
0.1 |
| 300PSI |
20BAR |
2KKPA |
2.0x |
0.01 |
0.001 |
0.01 |
0.1 |
1 |
|
0.1 |
0.001 |
1 |
| 500PSI |
30BAR |
3KKPA |
2.0x |
0.01 |
0.001 |
0.1 |
1 |
1 |
|
0.1 |
0.001 |
1 |
| 1KPSI |
70BAR |
7KKPA |
2.0x |
0.1 |
0.001 |
0.1 |
|
|
|
0.1 |
0.001 |
MPa |
| 2KPSI |
140BAR |
14KKPA |
2.0x |
0.1 |
0.01 |
0.1 |
|
|
|
1 |
0.01 |
0.001 |
| 3KPSI |
200BAR |
20KKPA |
1.5x |
0.1 |
0.01 |
0.1 |
|
|
|
1 |
0.01 |
0.001 |
| 5KPSI |
300BAR |
30KKPA |
1.5x |
0.1 |
0.01 |
1 |
|
|
|
1 |
0.01 |
0.001 |
| 10KPSI |
700BAR |
70KKPA |
1.5x |
1 |
0.01 |
|
|
|
|
1 |
0.01 |
0.001 |
Unneeded pressure units may be disabled via the RS-232 connector using ConFigXP software. • kPa models can display pressure in kPa and bar (or mbar) only. PSI and bar models can display all available units. • An XP2i will indicate pressure up to 10% above Range Pressure. Above 110%, the XP2i display will flash, indicating that the applied pressure exceeds the calibrated pressure range. If the calibrated pressure range is exceeded, the pressure displayed may not be accurate. • An Absolute version available only on XP2i gauges with ranges of 2000 PSI and higher. • MPa is available on -DD models only.
Options
Part No.
|
|
| -DD
|
Dual Line Display
Adds a second line to the display which can be set to indicate peaks, average, leak rate or differential pressure when connected to a second pressure gauge connected via an ordinary DB-9 null modem cable. Great for relief valve or leak testing. Cannot be ordered with the -DL DataLogging option.
|
| -DL
|
DataLoggerXP
Allows you to record up to 6000 measurements. The software application handles all of the transfer and saving of data to your computer's hard drive, while the XP2i once authorized, performs all of the actual data collection
|
| -B |
Absolute Pressure
Available for gauges rated at 2000 PSI or higher
|
| -F4
|
Panel Mounting
Includes a 4.5 inch gauge panel mount kit with a rear pressure port. Picture shows the kit installed. To replace a 6 inch or 8.5 inch panel mount gauge, you will need to order part number 2955 for the 6 inch flange adapter or part number 2956 for the 8.5 inch flange adapter. Installation instructions.
|
| -RP
|
Rear Port Fitting
The rear port fitting can be ordered separately but is included automatically with the panel mount flange option (F4)
|
| -O |
Cleaned for Oxygen Service
For applications requiring cleaning for oxygen
|
| -BSP |
G 1/4 B Parallel Pressure Fitting (Not available for PSI Models)
|
Software
Part No.
|
|
| -DL
|
DataLoggerXP
Allows you to record up to 6000 measurements. The software application handles all of the transfer and saving of data to your computer's hard drive, while the XP2i once authorized, performs all of the actual data collection
|
| ConFigXP
|
ConFigXP-Free
Allows you to customize your gauge by adding or disabling feature
|
| FastCalXP
|
FastCalXP calibration software that allows you to perform a 10 point up and down calibration in less than 3 minutes. See Demo |
|
LabVIEW driver
|
LabVIEW driver - Free
If using LabVIEW, the driver library implements the serial commands of the XP2i and provides parsing of the streaming data output
|
|
Crystal USB Driver 
|
Crystal USB driver - Free
Driver for part# 3313 USB-RS232 Adapter |
| Firmware upgrade |
XP2iUpdate - Free
Update your gauge to the latest Firmware to add new features like our fast Pressure Safety Valve (PSV) test mode
|
|
UnitCalc
|
UnitCalc - Free
Easy Pressure Unit Calculator. Displays all conversion factors and converted units at the same time. UnitCalc also calculates Head Correction, Gravity Correction and Air Density |
Accessories
Part No.
|
|
| 3193
|
Protective Boot - Ist Generation Gauge (Black Gasket around Gauge)
A Skydrol compatible elastomeric boot to increase shock resistance
|
| 3696
|
Protective Boot - 2nd Generation Gauge (No Black Gasket Around Gauge)
A Skydrol compatible elastomeric boot to increase shock resist |
3009
|
Hard Carrying Case (egg crate foam) |
2888
|
Weatherproof Hard Carrying Case (die-cut foam) |
| 2400 |
RS232 Cable (6ft)
XP2i can be calibrated via the keypad or the RS232 port or automatically using a calibration grade pressure controller
|
| 3456 |
Null Modem Adapter
A null modem for differential pressure testing between two XP2i
|
3313
 |
Adapter USB B receptacle to RS232 DB9M |
2955
|
Flange Adapter Kit
6 inch flange adapter kit requires a gauge with the -F4 panel mount option
|
| 2956 |
Flange Adapter Kit
8.5 inch flange adapter kit requires a gauge with the -F4 panel mount option |
| 2984 |
Universal AC Power Supply
Includes a set of plugs permitting use worldwide. In the event of a power failure, an XP2i powered by the AC power supply will automatically switch to its internal batteries
|
| Contact Crystal |
Diaphragm and sanitary seals |
|
|