Accuracy
20% to 100% of Full Scale: ±(0.1% of Reading).
0 to 20% of Full Scale: ±(0.02% of Full Scale).
Vacuum, for 2MPa (300 PSI, 20 bar, 20kg/cm2) and lower pressure gauges:
0 to -99.9 kPa (-14.5 PSIG): ±(0.25% of Full Scale), where F.S.: -99.9 kPa (-14.5 PSIG).
(Not recommended for continuous use at high vacuum.)
Pressure Ranges
(see table below for details)
Note:
Density of water conversion can be selected via the keypad: 4°C, 60°F or 20°C/68°F.
Temperature
Operating & Compensated: -10°C to 50°C (14°F to 122°F).
Storage: -40°C to 75°C (-40°F to 167°F).
Backlit Display
Description: 5 Full (seven segment) digits.
Display rate: 3 readings per second.
Numeral display height: 16.5mm (0.65") single line display or 14 mm (0.55") for main numerals of dual line display.
Sensor
Description: All welded stainless steel, with a permanent fill diaphragm seal (filled with Dow Corning 200).
Power
Battery: Three size AA (LR6) batteries.
Battery life: 1500 hours typical (alkaline battery).
Enclosure
Description: Aluminum alloy, sealed to IP67 per IEC 60529 ratings. (Excludes the WT Series)
Weight: 555g (19.6 oz.), including batteries.
|
Intrinsic Safety
Exia Intrinsically Safe, Class I, Division 1, Groups A, B, C & D, Temperature code T4

Other Certifications
DNV classification certifies all materials, components and systems relevant to the safe operation and quality of ships.

Connections

Pressure connection: 1/4 male NPT or G 1/4 B BSP.
Electrical connection: DB9, RS-232 (environmentally sealed).
Media Compatibility
Liquids and gases compatible with 316 stainless steel.
|