Differential Pressure Accuracy
20% to 100% of Full Scale ±(0.1% of Reading)
0% to 20% of Full Scale ±(0.02% of Full Scale)
Static line pressure effect is 0.0003 psi per lowest applied test pressure and can be eliminated by zeroing the gauge while the lowest test pressure is applied to left (+) and right (-) ports simultaneously.
Pressure Ranges
(see table below for details)
Note:
Density of water conversion can be selected via the keypad: 4°C, 60°F or 20°C /68°F.
Temperature
Operating & Compensated: -10°C to 50°C (14°F to 122°F).
Storage: -40°C to 75°C (-40°F to 167°F).
Backlit Display
Description: 5 Full (seven segment) digits.
Display rate: 3 readings per second.
Numeral display height: 16.5mm (0.65") single line display.
Power
Battery: Three size AA (LR6) batteries.
Battery life: 1500 hours typical (alkaline battery).
This unit is instrinsically safe only if powered by Rayovac Maximum Plus, Energizer E91, Energizer EN91, or Duracell MN 1500 batteries. Replace batteries only in non-hazardous locations. |
Enclosure
Description: 300 series SS sealed.
Dimensions: 33mm (1.3") D x 150 mm (5.9") W x 165 mm (6.5") L
Weight: 860g (1.9 lbs), including batteries.
Sensor
Sensor manifold includes a field replaceable burst disc to prevent over-pressure damage to the differential pressure sensor.
Max Static Line Pressure: 100 PSI
Intrinsic Safety
Exia Intrinsically Safe, Class I, Division 1, Groups A, B, C & D, Temperature code T4

Connections
Pressure Connections: 1/8" female NPT
Electrical Connection: DB9, RS-232 (environmentally sealed)
Do not use the RS-232 connector in a hazardous area.
Media Compatibility
Liquids and gases compatible with 300 Series stainless steel and buna-n (o-ring). |